MST-B series cabinet type pulsevacuum sterilizer uses saturated steam as sterilization medium, through pulsevacuum way, the saturated steam completely penetrates to the surface of the article, uses high temperature saturated steam under high pressure to kill all microorganisms and their spores, the sterilization effect is reliable, is the most effective physical sterilization method.It is widely used in pharmaceutical, biological engineering, laboratory and other fields to sterilize fabrics, instruments, lumen instruments, rubber, liquid and other items.


Technical parameters:
Design pressure: -0.1/0.3MPa;
Rated working pressure: 0.23MPa
  Ultimate vacuum: -0.096MPa 
Pressure display accuracy: 0.1kpa
Sterilization temperature: 105 ~ 136℃ 
Temperature display accuracy 0.1℃
Maximum temperature: 136℃ 
Temperature uniformity: ≤±0.5℃;
◆ Cavity material: SUS304 stainless steel

mode

volume

Chamber size

Overall dimensions

Weight

POWER

(L)

(Φ×L)

(L×W×H)

Kg

MST-45B

45

316×590

850×600×520

65

220V 50Hz  3.6Kw

MST-60B

60

380×500

985×600×1450

105

380V 50Hz  5.7Kw

MST-80B

80

380×700

985×600×1450

120

380V 50Hz  5.7Kw

MST-135B

135

500×700

1000×710×1550

200

380V 50Hz  9.35Kw

MST-185B

185

500×950

1250×850×1700

230

380V 50Hz  10.8Kw

MST-300B

300

630×1000

1350×950×1750

350

380V 50Hz  12Kw

Main performance and features:
Adopts microcomputer control technology, touch screen operation, the whole process (pulsevacuum, heating, sterilization, exhaust steam, vacuum drying) procedures automatically run.
Touch screen automatically displays operating status, temperature, pressure, time and other parameters and fault alarm information.
Has 4 sterilization procedures including instruments, dressings, luminal instruments, rubber, liquid and custom, BD test and vacuum test for users to choose.
Built-in vacuum pump, three times (can be set) pulsevacuum, vacuum degree up to -0.090mpa can be fully pumped out of dressings and class A cavity instruments.
Built-in high efficiency steam generator, can quickly produce constant pressure steam, do not pollute sterilized goods.
Vacuum drying system, fabric dryness less than 1%, instrument dryness less than 0.1%
0.22μm bacterial air filter.Ensure that the air breaking the vacuum is sterile air to avoid re-contamination
Optional configuration of micro printer, real-time sterilization process and fault alarm information.

Safety device:
Overtemperature protection device: when the temperature of the pot exceeds the set temperature, the system automatically cuts off the heating power and displays an alarm.
Door safety interlocking device: only when the door is closed in place, the program can start;The door will not open when there is dangerous pressure in the inner chamber.
Double overpressure protection: automatic pressure relief valve, pressure exceeds the design pressure, the safety valve opens to release pressure;Overpressure automatic protection control, pressure exceeds the set pressure, automatic exhaust pressure relief, display alarm.
Low water level alarm device: the water level is lower than the specified water level, automatically cut off the heating, display alarm